811401

ELISA Kit for Collagen Type XII (COL12) Rat

In Stock
*
Please Email support@aspirachemical.com to place your order.

In Stock
Products specifications
Species Rat
Format 96T
Sample Type Serum, plasma and other biological fluids.
Assay Length 4.5 h
Detection Range 0.313-20ng/mL. The standard curve concentrations used for the ELISA’s were 20ng/mL, 10ng/mL, 5ng/mL, 2.5ng/mL, 1.25ng/mL, 0.625ng/mL, 0.313ng/mL
Sensitivity 0.134ng/mL
Specificity This assay has high sensitivity and excellent specificity for detection of Collagen Type XII (COL12). No significant cross-reactivity or interference between Collagen Type XII (COL12) and analogues was observed.
Intra-Assay Precision <10%
Inter-Assay Precision <12%
Storage Temperature -20 °C
Shipping Conditions 2-8 °C
Handling Instruction The activity loss rate (stability) of this kit is less than 5% within the expiration date under appropriate storage conditions. To minimize extra influence on the stability, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. When not in use, kit components should be refrigerated. All reagents should be brought to room temperature before use. It is recommended that all standards, controls, and samples be run in duplicate. Do not mix or interchange different reagent lots from various kit lots.
Test Principle The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Collagen Type XII (COL12). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Collagen Type XII (COL12). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Collagen Type XII (COL12), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Collagen Type XII (COL12) in the samples is then determined by comparing the O.D. of the samples to the standard curve.
Notes All kits are custom made on-demand. Please allow 10-15 working days for production and 2-3 days for shipment.
Legal Information Marketed under license from USCN Life Science Inc.
ReagentsQuantity
Pre-coated, ready to use 96-well strip plate1
Standard2
Detection Reagent A1×120µL
Detection Reagent B1×120µL
TMB Substrate1×9mL
Wash Buffer (30 × concentrate)1×20mL
Plate sealer for 96 wells4
Standard Diluent1×20mL
Assay Diluent A1×12mL
Assay Diluent B1×12mL
Stop Solution1×6mL
Instruction manual1

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Collagen Type XII (COL12) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:81:16
serum(n=5)93-105%78-97%93-101%78-104%
EDTA plasma(n=5)94-102%80-105%83-93%83-104%
heparin plasma(n=5)97-105%79-98%96-105%87-97%

Matrices listed below were spiked with certain level of recombinant Collagen Type XII (COL12) and the recovery rates were calculated by comparing the measured value to the expected amount of Collagen Type XII (COL12) in samples.

MatrixRecovery range (%)Average(%)
serum(n=5)80-9686
EDTA plasma(n=5)94-10297
heparin plasma(n=5)95-10299
Products specifications
Species Rat
Format 96T
Sample Type Serum, plasma and other biological fluids.
Assay Length 4.5 h
Detection Range 0.313-20ng/mL. The standard curve concentrations used for the ELISA’s were 20ng/mL, 10ng/mL, 5ng/mL, 2.5ng/mL, 1.25ng/mL, 0.625ng/mL, 0.313ng/mL
Sensitivity 0.134ng/mL
Specificity This assay has high sensitivity and excellent specificity for detection of Collagen Type XII (COL12). No significant cross-reactivity or interference between Collagen Type XII (COL12) and analogues was observed.
Intra-Assay Precision <10%
Inter-Assay Precision <12%
Storage Temperature -20 °C
Shipping Conditions 2-8 °C
Handling Instruction The activity loss rate (stability) of this kit is less than 5% within the expiration date under appropriate storage conditions. To minimize extra influence on the stability, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. When not in use, kit components should be refrigerated. All reagents should be brought to room temperature before use. It is recommended that all standards, controls, and samples be run in duplicate. Do not mix or interchange different reagent lots from various kit lots.
Test Principle The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Collagen Type XII (COL12). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Collagen Type XII (COL12). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Collagen Type XII (COL12), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Collagen Type XII (COL12) in the samples is then determined by comparing the O.D. of the samples to the standard curve.
Notes All kits are custom made on-demand. Please allow 10-15 working days for production and 2-3 days for shipment.
Legal Information Marketed under license from USCN Life Science Inc.