810473

ELISA Kit for Anti-Mullerian Hormone (AMH) Swine

In Stock
*
Please Email support@aspirachemical.com to place your order.

In Stock
Products specifications
Species Swine
Format 96T
Sample Type Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids.
Assay Length 2.5 h
Detection Range 370.37-30000pg/mL. The standard curve concentrations used for the ELISA’s were 30000pg/mL, 10000pg/mL, 3333.33pg/mL, 1111.11pg/mL, 370.37pg/mL
Sensitivity 135.8pg/mL
Specificity This assay has high sensitivity and excellent specificity for detection of Anti-Mullerian Hormone (AMH). No significant cross-reactivity or interference between Anti-Mullerian Hormone (AMH) and analogues was observed.
Intra-Assay Precision <10%
Inter-Assay Precision <12%
Storage Temperature -20 °C
Shipping Conditions 2-8 °C
Handling Instruction The activity loss rate (stability) of this kit is less than 5% within the expiration date under appropriate storage conditions. To minimize extra influence on the stability, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. When not in use, kit components should be refrigerated. All reagents should be brought to room temperature before use. It is recommended that all standards, controls, and samples be run in duplicate. Do not mix or interchange different reagent lots from various kit lots.
Test Principle This assay employs the competitive inhibition enzyme immunoassay technique. An antibody specific for Anti-Mullerian Hormone (AMH) has been pre-coated onto a microplate. A competitive inhibition reaction is launched between Horseradish Peroxidase (HRP) labeled Anti-Mullerian Hormone (AMH) and unlabeled Anti-Mullerian Hormone (AMH) (Standards or samples) with the pre-coated antibody specific for Anti-Mullerian Hormone (AMH). After incubation the unbound conjugate is washed off. The amount of bound HRP conjugate is reverse proportional to the concentration of Anti-Mullerian Hormone (AMH) in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of Anti-Mullerian Hormone (AMH) in the sample.
Notes All kits are custom made on-demand. Please allow 10-15 working days for production and 2-3 days for shipment.
Legal Information Marketed under license from USCN Life Science Inc.
ReagentsQuantity
Pre-coated, ready to use 96-well strip plate1
Standard2
Detection Reagent A1×120µL
Detection Reagent B1×120µL
TMB Substrate1×9mL
Wash Buffer (30 × concentrate)1×20mL
Plate sealer for 96 wells4
Standard Diluent1×20mL
Assay Diluent A1×12mL
Assay Diluent B1×12mL
Stop Solution1×6mL
Instruction manual1

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Anti-Mullerian Hormone (AMH) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:81:16
serum(n=5)78-98%79-89%84-96%90-99%
EDTA plasma(n=5)93-102%84-95%96-104%92-105%
heparin plasma(n=5)81-92%90-104%80-93%78-102%

Matrices listed below were spiked with certain level of recombinant Anti-Mullerian Hormone (AMH) and the recovery rates were calculated by comparing the measured value to the expected amount of Anti-Mullerian Hormone (AMH) in samples.

MatrixRecovery range (%)Average(%)
serum(n=5)88-10491
EDTA plasma(n=5)97-105102
heparin plasma(n=5)86-9995
Products specifications
Species Swine
Format 96T
Sample Type Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids.
Assay Length 2.5 h
Detection Range 370.37-30000pg/mL. The standard curve concentrations used for the ELISA’s were 30000pg/mL, 10000pg/mL, 3333.33pg/mL, 1111.11pg/mL, 370.37pg/mL
Sensitivity 135.8pg/mL
Specificity This assay has high sensitivity and excellent specificity for detection of Anti-Mullerian Hormone (AMH). No significant cross-reactivity or interference between Anti-Mullerian Hormone (AMH) and analogues was observed.
Intra-Assay Precision <10%
Inter-Assay Precision <12%
Storage Temperature -20 °C
Shipping Conditions 2-8 °C
Handling Instruction The activity loss rate (stability) of this kit is less than 5% within the expiration date under appropriate storage conditions. To minimize extra influence on the stability, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. When not in use, kit components should be refrigerated. All reagents should be brought to room temperature before use. It is recommended that all standards, controls, and samples be run in duplicate. Do not mix or interchange different reagent lots from various kit lots.
Test Principle This assay employs the competitive inhibition enzyme immunoassay technique. An antibody specific for Anti-Mullerian Hormone (AMH) has been pre-coated onto a microplate. A competitive inhibition reaction is launched between Horseradish Peroxidase (HRP) labeled Anti-Mullerian Hormone (AMH) and unlabeled Anti-Mullerian Hormone (AMH) (Standards or samples) with the pre-coated antibody specific for Anti-Mullerian Hormone (AMH). After incubation the unbound conjugate is washed off. The amount of bound HRP conjugate is reverse proportional to the concentration of Anti-Mullerian Hormone (AMH) in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of Anti-Mullerian Hormone (AMH) in the sample.
Notes All kits are custom made on-demand. Please allow 10-15 working days for production and 2-3 days for shipment.
Legal Information Marketed under license from USCN Life Science Inc.