Products specifications
Species
|
Rat
|
Format
|
96T
|
Sample Type
|
Serum, plasma and other biological fluids.
|
Assay Length
|
2.5 h
|
Detection Range
|
12.35-1000pg/mL. The standard curve concentrations used for the ELISA’s were 1000pg/mL, 333.33pg/mL, 111.11pg/mL, 37.04pg/mL, 12.35pg/mL
|
Sensitivity
|
4.39pg/mL
|
Specificity
|
This assay has high sensitivity and excellent specificity for detection of Adrenocorticotropic Hormone (ACTH).
No significant cross-reactivity or interference between Adrenocorticotropic Hormone (ACTH) and analogues was observed.
|
Intra-Assay Precision
|
<10%
|
Inter-Assay Precision
|
<12%
|
Storage Temperature
|
-20 °C
|
Shipping Conditions
|
2-8 °C
|
Handling Instruction
|
The activity loss rate (stability) of this kit is less than 5% within the expiration date under appropriate storage conditions. To minimize extra influence on the stability, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. When not in use, kit components should be refrigerated. All reagents should be brought to room temperature before use. It is recommended that all standards, controls, and samples be run in duplicate. Do not mix or interchange different reagent lots from various kit lots.
|
Notes
|
All kits are custom made on-demand. Please allow 10-15 working days for production and 2-3 days for shipment.
|
Legal Information
|
Marketed under license from USCN Life Science Inc.
|
Reagents | Quantity |
Pre-coated, ready to use 96-well strip plate | 1 |
Standard | 2 |
Detection Reagent A | 1×120µL |
Detection Reagent B | 1×120µL |
TMB Substrate | 1×9mL |
Wash Buffer (30 × concentrate) | 1×20mL |
Plate sealer for 96 wells | 4 |
Standard Diluent | 1×20mL |
Assay Diluent A | 1×12mL |
Assay Diluent B | 1×12mL |
Stop Solution | 1×6mL |
Instruction manual | 1 |
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Adrenocorticotropic Hormone (ACTH) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample | 1:2 | 1:4 | 1:8 | 1:16 |
serum(n=5) | 95-105% | 90-102% | 95-103% | 90-97% |
EDTA plasma(n=5) | 90-101% | 88-104% | 89-103% | 83-91% |
heparin plasma(n=5) | 88-103% | 94-101% | 88-102% | 87-101% |
Matrices listed below were spiked with certain level of recombinant Adrenocorticotropic Hormone (ACTH) and the recovery rates were calculated by comparing the measured value to the expected amount of Adrenocorticotropic Hormone (ACTH) in samples.
Matrix | Recovery range (%) | Average(%) |
serum(n=5) | 85-93 | 88 |
EDTA plasma(n=5) | 79-101 | 84 |
heparin plasma(n=5) | 80-94 | 91 |